Sper-scientific 300037 Refractometer - Programmable Manual do Utilizador Página 6

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 8
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 5
6
played.
 Press the READ button and the last digit flashes.
 Use the and the buttons to set the high
range limit.
 Press the READ button to activate the function.
 Press the FUNC button as needed to return to
the measurement screen.
Deactivating High Range Alert
 Press the FUNC button as needed until HLn
appears.
 Press the READ button and the last digit flashes.
 Press the FUNC button as needed to return to
the measurement screen.
Setting Low Range Alert
 With the unit on, press the FUNC button as need-
ed until LLYand the flashing L-Limit icon appear.
 Press the READ button and the last digit flashes.
 Use the and buttons to set the low range
limit.
 Press the READ button activate the function.
 Press the FUNC button as needed to return to
the measurement screen.
Deactivating Low Range Alert
 Press the FUNC button as needed until LLn
appears.
 Press the READ button and the last digit flashes.
 Press the FUNC button as needed to return to
the measurement screen.
6. ERRATIC READINGS
A. If the readings are significantly above or below the
expected results, consider the following:
B. Sample contains more than one dissolved solid.
For example: if the sample contains both salt and sugar, a Brix
Refractometer will not display the true sugar content because the
salt content will skew the Brix measurement.
Vista de página 5
1 2 3 4 5 6 7 8

Comentários a estes Manuais

Sem comentários