7. NOTES
Emissivity is the ability of an object to emit or absorb en-
ergy. Your IR Gun measures emitted energy using a
fixed emissivity value of 0.95 (which covers about 90%
of typical applications). When measuring highly reflective
surfaces, apply masking tape to the surface or a paint
that has a 0.95 emissivity. If the object you are measur-
ing is covered with frost, clear the frost to expose the
object's surface before taking the measurement.
Your IR Gun has an 8:1 Distance-to-Spot ratio.
The target must be larger than the IR Gun's spot size.
(The smaller the target, the closer you must be to it.)
Your reading may fluctuate if the IR Gun is moved
during measurement.
Allow the IR Gun to adjust to the ambient temperature
for 1/2 hour before use for the most accurate results.
This IR Gun is not intended for medical evaluations.
Protect the IR Gun from water, physical or electrical
shock, dust, and extreme environments.
Contains a laser pointer. Keep out of reach of children.