III. Operating Instructions
Precautions and Maintenance
Do not store or operate the instrument at
high temperatures or humidity.
Keep the microphone dry and avoid se-
vere shock or vibration.
Wind blowing across the microphone can
cause measurement errors, these effects
are reduced by using the windscreen.
Remove the batteries if the unit is not in
use for a long period of time.
Wipe the case with a dry cloth, do not
use abrasives or solvents.
Calibration
Calibrate the meter before first use or
when the reading appears inaccurate.
If you require a certificate of calibration
traceable to NIST, the unit may be sent to
Sper Scientific for servicing (fee applies).
There are two methods of calibration:
1. Internal Electrical Calibration
With the unit on, lightly press the
Battery Cover to slide it down and off.
Move the Measurement Slide Switch
down. This activates the built-in
oscillator (94dB 1 kHz, sine wave) for