114. Press ↑ to increase the value by 1. Hold down ↑ to continually increase the value. Press ↓ to decrease the value by 1. Hold down ↓ to continua
11Restore Factory Settings1. Press SETUP to enter the Setup Mode.2. Press ↑ or ↓ to select the Restore Factory Settings screen (see Fig. A). 3. Pres
13Set MTC (Manual Temperature Compensation) Temperature CoefcientNote...Only perform this setting when ATC Temperature Probe is not available.1. P
13View the Probe Data1. Press SETUP once to enter the Set up mode.2. Press ↑ or ↓ to select the View Probe screen (see Fig. A).3. Press SETUP once t
14 15Set Barometric Pressure Parameters1. Press SETUP to enter the Setup Mode. 2. Press ↑ or ↓ until the Bar. Pressure screen is seen (see Fig. A).3.
15Set RH Parameters1. Press SETUP to enter the Setup mode.2. Press ↑ or ↓ to display the Set RH screen (see Fig. A)3. Press SETUP to enter the Set RH
17Set Salinity Compensation1. Press RANGE to select mg/L (ppm) measurement scale.2. Press SETUP to enter the Setup Mode.3. Press ↑ or ↓ to display the
17Set % Offset1. Press RANGE to select % measuring scale.2. Press SETUP to enter the Setup Mode.3. Press ↑ or ↓ to display Set % Offset screen (see Fi
18 19CALIBRATIONNote...While performing calibration, if the measurement value reads from 10% to 50% the display will show “Error” as in Fig. A. Check
19100% CalibrationNote...Only when the measurement value is above 50% can the meter be calibrated at 100%. All other keys except POWER ON/OFF and HOLD
2 3AquaShock® DO Kit 850046KCopyright ©2014 by Sper ScienticALL RIGHTS RESERVEDPrinted in the USAThe contents of this manual may not be reproduced or
20 21Fig. A Fig. BFig. C Fig. D0% CalibrationNote...The measurement value must read below 10% to calibrate the meter at 0%. All other keys except POWE
21Calibrate dissolved oxygen in mg/L (ppm) modeNote...mg/L (ppm) calibration can only be performed after 100% calibration has been completed.1. Clean
22 23View Calibration Points1. Press RANGE chose the scale you want view.2. Press SETUP to enter the Setup mode.3. Press ↑ or ↓ to select the View Cal
23Fig. E Fig. FAuto RangingDuring measurement and calibration this meter will automatically switch to the correct range.MEASUREMENT PROCEDURESNote...B
24 25Scale Selection1. Press RANGE to switch between the % of Saturation and Milligrams per Liter (mg/L) scales (see Fig. A and Fig. B).Backlight1.
25Recall Memory1. Press RANGE to select the scale for recall. 2. Press MR to enter the Recall mode and view the last saved reading (see Fig. A and F
26 27Hold1. Press HOLD to hold the value (see Fig. A).2. Press HOLD to release the hold value.Note…When in Hold Mode, all other functions will be disa
27Fig. A Fig. BFig. CFig. DFig. E Fig. FFig. G Fig. HMode mg/L(ppm) % Saturation TemperatureRange 0.00 ~ 19.99 mg/L (ppm)0.0 ~ 199.9% 0 ~ 50°C 32 ~ 12
28 PBWARRANTY Sper Scientic warrants this product against defects in materials and workmanship for a period of ve (5) years from the date of purchas
3INTRODUCTION ...4FEATURES...5MATERIALS SUPPLIED...
4 5INTRODUCTIONThe Sper Scientic AquaShock® DO Meter is the most waterproof, rugged and reliable DO meter on the market. It is fully waterproof, incl
5FEATURES• Floats• Shockproof• IP67 Waterproof• Protective soft grip outer layer• Rechargeable Lithium Ion battery• Multi-line display LCD screen• Rea
6 7LCD DISPLAYName Description1 Ready Displayed when measuring results are stable2 Scale Unit Current measurement scale unit3 Measuring Results Curre
7Battery voltage is more than 7.7V shows full.Battery Power Icon:Battery voltage between 7.2V and 7.7V shows 2 cells.Battery voltage between 6.6V and
8 9DO Meter13574682KEYPADName 1st Function 2nd Function1 Power/Backlight/Low BatteryPower on/off Backlight on/off2 Setup Setting parameters Conrm set
9SETUP MODEThe Setup Mode allows you to customize the following meter preferences and defaults: • Temperature Scale• Clock Setting• Restore Factory D