- 11 -
This procedure requires an optional 2-
Point Acoustical Calibrator (850016):
1. Using the directions that accompany
the Acoustical Calibrator, ensure that
the calibrator is operational and is
equipped with the ½” microphone
adapter.
2. Insert meter’s sound probe into the
calibrator.
3. Turn the calibrator on and select 94
dB.
4. Turn the meter on and use the
FUNCTION button to select the Sound
–Level meter function.
5. Press HOLD and MAX/MIN
simultaneously for 3 seconds and
release.
SOUND CALIBRATION