INTRODUCTION
This simple and easy to use waterproof
pH Pen measures from 2~12 pH.
DISPLAY / FACEPLATE
CALIBRATION
 Calibrate routinely before use, an-
ytime the readings seem erratic, and
whenever the batteries are replaced.
 Remove the protective cap.
 Clean the electrode with de-ionized/
distilled water. Note: Crystals may
form in the cap and these can be
soaked or rinsed away. Do not touch
or rub the glass bulb.
 Before first use or after extended
storage, soak the electrode in
distilled water for 10~30 minutes.
 Press >1 sec to turn on the unit.
 For 1-point calibration, dip/stir the
electrode into pH 7 buffer solution
and press the button. If the
probe is damaged or cannot detect
the buffer, the unit automatically exits
calibration mode after 10 seconds.
 During calibration, "CA" flashes on
the display.
 Then, "CAL" and the pH value are
shown.
 "SA" (save) is displayed as the
calibration point is saved.
 "En" (end) displays when the
calibration ends.
 For 2 or 3 point calibration, start with
pH 7, then repeat the above steps for
pH 4 and/or pH 10.
 For highest accuracy, calibrate at the
sample temperature.
 Repeat the above procedures until
readings stabilize.
 NOTE: If the buffer value is not
standard (for example 4.2), press
to adjust the displayed value.
"SA" (save) is displayed, followed by
"En" (end).
MEASUREMENT
 Calibrate the pen as needed.
 Remove the protective cap and clean
the electrode with de-ionized/distilled
water. Do not touch the glass bulb.
 Press >1 sec to turn on the unit.
 Dip/stir the electrode into the sample.
 The pH value is displayed.
 "- -
-
" is displayed when the pH
sample is out of range.
 During measurement, press to
freeze the reading on the LCD.
"Hold" is displayed. Press again to
return to normal measurement.
 The pH Pen automatically shuts off
after 20 minutes of inactivity. To
disable this feature, start with the unit
off, then simultaneously press the
and buttons. Release both buttons
when "n" is displayed. The unit
returns to measurement mode.
 After use, turn off the unit and rinse
the electrode.
 Protect the electrode's lifespan by
adding a few drops of KCL (potassium
chloride) solution into the protective
cap before storage.
 Always keep the protective cap on
when the unit is not in use.
Rev. 3/29/2011
Pull to
Remove the
Protective Cap
Electrode
FIRST USE: If the
unit fails to turn on,
check that the
batteries are working
and properly
installed.
HoldCal
7.0
ATC
87.5
C
F
pH Value
Symbol will
flash when
battery is low
Hold
button
Calibration
button
On/Off
button
HLD
CAL
Vista de página 0
1 2

Resumo do Conteúdo

Página 1 - HoldCal

INTRODUCTION This simple and easy to use waterproof pH Pen measures from 2~12 pH. DISPLAY / FACEPLATE CALIBRATION  Calibrate routinely before us

Página 2 - SPER SCIENTIFIC

LARGE DISPLAY PH PEN 850050 WARRANTY: Sper Scientific warrants this product against defects in materials a