- 7 -
4. ADVANCED AJUSTMENT PROCEDURES
Press the SET button (8) for at least two seconds to enter the advanced
adjustment procedures. Press the ESC (3) button to exit.
4-A. Check Remaining Available of Data Points
Press the SET button (8) at least 2 seconds until the lower display shows:
"XXXXX" is the number of free data points.
4-B. Clear Memory
Press the SET (8) button as needed to reach the
"clear memory" screen. The number of saved data
points is displayed.
To delete saved data from memory, press the ENTER (4) button once,
then press ENTER (4) button again to confirm your action. The display
shows: "0 Ok ESC:Quit".
If you DO NOT want to clear the memory, press the ESC (3) button to
exit, or press the SET (8) button to advance to the next option.
4-C. Date/Time Setting
Press the SET button (8) as needed to reach
the "date/time set" screen.
Use the UP (5), DOWN (6), and ENTER (4) buttons to select the
year, month, date, hours, minutes and seconds.
Press the ENTER (4) button to save the settings.
Press the SET (8) button to advance to the next option, or press the ESC
(3) button to exit.
4-D. Sample Time Setting
Press the SET (8) button as needed to reach
the "sample time" screen.
Use the (5), (6), and ENTER (4) buttons to select the sample time
(hour-min-sec).
Press the ENTER (4) button to save the settings.
Press the SET (8) button to advance to the next option, or press the ESC
(3) button to exit.
00:00:02 Sample Time
h-m-s ^, v ENTER()
06:01:03 Date/Time Set
y-m-d ^, v ENTER()
XXXXX Memory Space
Before executing the advanced adjustment procedures,
exit the HOLD and RECORD functions.
Clear Memory
Esc:N Enter:Y
107