- 10 - 7. BATTERY REPLACEMENT When the left corner of the LCD displays the low battery icon install 4 new AA 1.5V batteries. Slide the BATTERY COV
- 11 - Prepare a 1.413 µS calibration solution, or any standard conductivity solution. Press the POWER (2) button to turn the me-ter on. Sele
- 12 - Rev. 01/23/2012 WARRANTY Sper Scientific warrants this product against defects in materials and workmanship for a period of five (5) years
TABLE OF CONTENTS 1. INTRODUCTION... 3 2. PANEL DESCRIPTION ...
- 3 - 1. INTRODUCTION Automatically records up to 16,000 (!) data points in the field along with time and date. These can later be downloaded to a co
- 4 - 2. PANEL DESCRIPTION 1 DISPLAY 2 POWER button 3 HOLD / ESC button 4 REC / ENTER button 5 ▲(UP) button 6 RANGE / ▼(DOWN) button 7 CLOCK/ SEN
- 5 - 3. MEASUREMENT PROCEDURES: See page 4 for button locations. Factory Preset Meter Defaults: Conductivity (uS, mS), °C, Temperature Com-pensatio
- 6 - 3-D. Maximum / Minimum To record the maximum and minimum readings: Press the REC (4) button once. "REC" appears on the LCD. P
- 7 - 4-A. Check Remaining Number of Data Points Press the SET (8) button at least 2 seconds until the lower display shows: "XXXXX" is the
- 8 - 4-F. Temperature Unit Default Setting Press the SET (8) button as needed to reach the "temp. unit" screen. Use the ▲ (5) and
- 9 - Data 1 Data X+1 Data Z To Data X To Data Y